1. Donwload the Software copy to a USB and connect it to the unit.
2. Press USB update Video Instructions Available
Update via usb
1. PCB Board output K16 failure.
2. New upgrade available
1. Assure the CFL step was not Skipped.(Indicated on the test report too)
2. Assure injector Warm-up was Not Skipped. (Indicated on the test report too)
3. Check In-Line filters and replace periodically
4. Replace the Calibration Oil with fresh/new one.(Check the timers)
5. Assure Not to use the flush adpater for measurement. Only Use of RSP or iPSC (Spray Chamber).
6. Assure the Pressure is correct by doing a verification (MVC1) using a Master hight pressure Gauge.
Please make the CFL 1/2/3 before test. Do not skip this step.
1. Check the connection of RSP adapter
2. Check the cable of RSP (cut/ damaged cable)
Important Do not skip the CFL tests. It is critical for the measurement of R2LC
1. Possible PCB failure
2. Connect a Solenoid /Coil injector to the Injector cable
3. Perform a machine Diagnostic test in the Diagnosis Screen/p>
1. Check the Injector wire connector used to be the correct one
2. Check if Proper wire extension is used for correct polarity (SVI.01 / BPI.01)
1. Clean all in line filters
2. Use a high pressure gauge and run the MVC1 in order to verify the pressure.
3. Assure the Calibration Oil is not overused and replaced ordinary, otherwise replace it.
1. Reinstall/install the application
2. Press “forget WiFi” on the connected WiFi network and then connect again as a new connection. Finally you will need to restart the unit.
Needs to Install New Printhand Application / update
Needs to Install New DropBox Application / update
test
While Copying the ACF file from the USB Memory Stick there was a failure due to the following reasons:
1. The USB contains two (2) files with the same extension, therefore cannot do it.
2. The .acf extension should be written in small letters and NOT Capitals.
Instructions available
Send email to Support with a picture or the information listed in the About screen in the machine software, so we can issue and Activation file for the unit. Functions and Operations are not limited when this message is active.
CZ R Unit Activation Procedure
Database is continiosuly being updated and upgraded with new injector profiles. Await the next updates or use the All Units – Generic Profile provided to make your own Database by comparing with new one.